PTFE packing standard, Graphite packing for high temperature on request
Available with 2 way remotely mounted and separately mounted valve.
Available with Pipe to Pipe, Pipe to Flange, Flange to Flange, Remote,
Remote mounting Base vent, Base, Direct and Integral mounting.
Convenient method of blocking, bleeding and calibrating pressure instruments.
Designed for connection system impulse line & transfer
Combined the function of a tee, calibration valve, isolation valve, all tubing and fittings in a single valve configuration.
Standard port sizes up to 1/2” NPT/BSP/BSPT
Operating Parameters
Pressure Rating : upto 6000 psig (408 bar) at CWP
Temperature Rating
-54ºC to 232ºC with PTFE packing
upto 428ºC with optional graphite packing
Applications
Pressure and differential pressure instrumentations.
Testing
Standard Test : Each Valve is factory tested with nitrogen at 1000 psig (69 bar) for leakage at the seat and packing, the maximum allowable leak rate of 0.1 sccm
Optional Hydrostatic test : This test is performed with de-ionised water at 1.5 time the working pressure